Change the Oil Mist Filter

SCD and NCD User Manual
The oil mist filter on the RV5 pump has two components: the
charcoal odor filter and the oil coalescing filter element. To
replace the filters, disassemble the oil mist filter assembly with
the 4 mm long-handled hex wrench (provided). The smaller
charcoal odor filter sits on top of the larger oil coalescing filter
element. While it is recommended to replace the oil coalescing
filter element after 90 days of continuous use, replacement of
the charcoal odor filter is optional. After replacing the filter,
re-assemble the filter assembly and attach it to the pump flange.
Reset the EMF counter.