Detector Stability and Response

SCD and NCD User Manual
The time required for system stabilization varies depending on
the application, system cleanliness, the presence of active sites,
and other factors.
• When starting an existing system, typically wait at least
10 minutes before using the system to collect data.
• A new burner or a new set of ceramic tubes may take up to
24 hours to condition. Set the detector to the operating
conditions, and monitor the baseline until the baseline
becomes stable enough for your application.
4
Operation
51