7
Performance Verification

Prepare sample vials

WA RNING
108
Table 13
Recommended parts for checkout by inlet type (continued)
Recommended part for checkout
Syringe, 5-µL on-column
0.32-mm needle for 5-µL syringe
7693A ALS: Needle support insert, COC
7683B ALS: Needle support assembly for
0.25/0.32 mm injections
Insert, fused silica, 0.32-mm id
Performance verification requires a 1 uL injection. The
checkout standards for SCD and NCD come with 3 ampuoles.
When handling checkout standards, always follow the handling
precautions recommended with its packaging.
1
Open the sample box.
2
Snap the top off of one checkout sample ampoule.
3
Transfer the contents to a 2-mL ALS sample vial and cap the
vial.
Part number
5182-0836
5182-0831
G4513-40529
G2913-60977
19245-20525
SCD and NCD User Manual