Add Vacuum Pump Oil

WA RNING
WA RNING
C A U T I O N
SCD and NCD User Manual
Add pump oil when the pump oil level is low.
Materials needed
• Funnel (9301-6461)
• 5-mm Allen wrench (8710-1838)
• Gloves, chemical resistant, clean, lint free (9300-1751)
• Oil, synthetic, Mobile 1 (G6600-85001)
• Safety glasses (goggles)
Never add pump oil while the pump is on.
The fill cap and pump may be dangerously hot. Check that the fill
cup and pump are cool before you touch them.
Use only synthetic 10W30 oil, such as Mobil 1. Any other oil can
substantially reduce pump life and invalidates the pump warranty.
Procedure
1
Shut down the detector and wait for the pump to turn off.
See
"Shutdown"
on page 56.
2
Turn off the detector and unplug the pump power cord at the
pump.
5
Maintenance
81