6
Troubleshooting

Coking

102
Contamination from some sample matrices can reduce
sensitivity. For example, crude oils containing volatile metal
complexes may contaminate the ceramic tubes. In addition,
incomplete combustion of certain hydrocarbon-containing
compounds leaves behind coke deposits on the ceramic tubes.
Coke deposits may be removed from the burner by reducing the
hydrogen flow rate.