Agilent Technologies 8355 Sulfur User Manual page 115

Chemiluminescence detectors
Hide thumbs Also See for 8355 Sulfur:
Table of Contents

Advertisement

SCD and NCD User Manual
However, a new burner (or a burner with a new quartz tube)
can have a very high baseline after first ignition. In this case,
the baseline should gradually decrease, depending on burner
cleanliness. Noise will also greatly decrease over time. For a
well-stabilized system, noise measured by Agilent OpenLAB
CDS should be approximately 4 display units or less.
Checkout can continue before the baseline becomes
completely stable.
6
Create or load a method with the parameter values listed in
Table
15.
Table 15
NCD Checkout conditions
Column and sample
Type
Sample
Column flow
Column mode
Split/splitless inlet
Temperature
Mode
Purge flow
Purge time
Septum purge
Gas saver
Multimode inlet
Mode
Inlet temperature
Initial time
Purge time
Purge flow
Septum purge
Gas saver
Cool on-column inlet
Temperature
Performance Verification
HP-5, 30 m × 0.32 mm × 0.25 µm
(19091J-413)
NCD checkout 5190-7002
2.2 mL/min
Constant flow
250 °C
Splitless
80 mL/min
0.8 min
3 mL/min
Off
Splitless
250 °C
0 min
0.8 min
80 mL/min
3 mL/min
Off
Oven Track
7
115