Safety and Regulatory Information

SCD and NCD User Manual
This instrument has been designed and tested in accordance
with IEC Publication 61010-1 Safety Requirements for Electrical
equipment for Measurement, Control, & Laboratory Use, and
has been supplied in a safe condition. The instruction
documentation contains information and warnings which must
be followed by the user to ensure safe operation and to maintain
the instrument in a safe condition.
Refer to the 7890B GC Safety Manual, available on the GC and
GC/MS User Manuals & Tools DVD, for general safety and
regulatory information for the 7890B GC. The sections below
provide information specific to the 8355 SCD and 8255 NCD.
The Agilent 8355 SCD and 8255 NCD conform to the following
safety standards:
• Canadian Standards Association (CSA): C22.2 No. 61010-1
• CSA/Nationally Recognized Test Laboratory (NRTL):
ANSI/UL 61010-1
• International Electrotechnical Commission (IEC): 61010–1,
61010-2-010
• EuroNorm (EN): 61010–1
This is a safety Class I, Pollution Degree 2, Installation
Category II product. It must be wired to a mains supply with a
protective earthing ground incorporated into the power cord.
Any interruption of the protective conductor, inside or outside
the equipment, is likely to make the instrument dangerous.
Intentional interruption is prohibited.
If this instrument is used in a manner not specified by Agilent,
the protection provided by the instrument may be impaired.
The Agilent 8355 SCD and 8255 NCD conform to the following
regulations on Electromagnetic Compatibility (EMC) and Radio
Frequency Interference (RFI):
• CISPR 11/EN 55011: Group 1, Class A
• IEC/EN 61326
• AUS/NZ
This ISM device complies with Canadian ICES-001. Cet appareil
ISM est conforme a la norme NMB—001 du Canada.