Barco MCM-100S User Manual page 127

Table of Contents

Advertisement

Image 12-30
3. In the Dynacolor Set dialog box, click Calibration.
A red full screen test pattern is displayed.
4. Use a spectrometer to measure the exact x and y value and brightness of the displayed color and press ENTER to proceed.
5. In the Calibration dialog box , fill in the measured x and y coordinates and the Y (i.e. brightness) value.
Image 12-31
Note: The Y value is transformed to a relative L value, which implies that the physical unit is irrelevant.
6. Click Proceed.
A Green test pattern is displayed
7. Proceed in the same way for Green, Blue and White as for Red.
8. The values are finally updated in the Measured values of the Dynacolor set.
9. Click Default desired to copy the Measured values for Red, Green and Blue and White to the Desired values and to calculate
the secondary colors Cyan, Yellow and Magenta.
R5905187 MCM-100S 05/06/2012
12. Expert operation
123