Page 2
BarcoView Th. Sevenslaan 106, B-8500 Kortrijk Phone: +32 56.23.32.44 Fax: +32 56.23.33.74 E-mail: sales.medical.barcoview@barco.com Visit us at the web: www.barcomedical.com Printed in Belgium...
Page 3
Federal Communications Commission (FCC Statement) This equipment has been tested and found to comply with the limits for a class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
1. ABOUT THIS MANUAL 1.1 Objective This manual is dedicated to the Barco MGP LCD projector and describes the way the projector should be operated and the use of the software menus . In the practice the MGP is to be implemented in a total medical visualization system . For this reason a chapter is inserted giving a brief introduction to medical systems and in particular the Barco’s DICOM THEATRE...
Page 10
1. About this manual R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
1 safety manual 2.2 Projector Packaging Way of Packaging The projector is packed in a carton box. To provide protection during transportation, the projector is surrounded with foam. The package is secured with banding and fastening clips. To unpack 1. Release the fastening clips. (image 2-1) 2.
Save the original shipping carton and packing material, they will be necessary if you ever have to ship your projector. For maximum protection, repack your projector as it was originally packed at the factory. Save the original shipping carton and packing material, they will be necessary if you ever have to ship your projector.
Page 13
2. Packaging and Dimensions Length with different lenses, see table Cold air in Hot air out Image 2-3 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 14
Lens length of projector 545mm SVD(2-2.5:1) 600mm QVD(1.3-1.8:1) QVD(1.9-2.6:1) 590mm QVD(3.0-6.0:1) 620mm QVD(0.85:1) 545mm QVD(7.0:1) 540mm combinable with cable Remarks basket Length with cable basket = 565mm Length with cable basket = 560mm R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
It is advised to clean the dustfilters on a regular basis and to replace them at any lamp change. Barco reserves itself the right to refuse warranty replacement of consumables if they have been used in a projector with dirty airfilters.
Image size The projector is designed for projecting an image size with a screenwidth from 1.00m (3.3ft) to 6.00m (19.7ft) with an aspect ratio of 4 to 3. 3.3 Projector configurations The different configurations The MGP can be installed in either a single or dual projector configuration..
Page 17
Lens range to be used in combination with the MGP Ceiling mount Kit for front projection: SVD (2.0 - 2.5) Only on axis projection since vertical shift is impossible (vertical shift becomes horizontal shift). Rear projection is also possible within a dedicated rear structure. For the landscape viewing mode one can define the standard projector configurations: Rear/Ceiling Rear/Table...
Lens installation • Removing the lens • Cleaning the lens 3.4.1 Lenses Available lenses The following lenses are available, or will become available (contact a BARCO service center) as an option : Lenses QVD(0.85:1) QVD(1.3-1.8:1) QVD(1.9-2.6:1) QVD(3.0-6.0:1) QVD(7:1) SVD(2.0-2.5:1) The QVD (0.85:1) is to be used in ON Axis configuration only.
How to remove the lens ? 1. Slide the lens door to the left. 2. Unlock the lens by pulling the handle located on the right side of the projector (image 3-8) 3. Remove the lens out of its housing...
3. Installation Guidelines AUTION Never transport the projector with the lens mounted on it ! Always remove the lens before transporting the projector. 3.4.5 Cleaning the lens To minimize the possibility of damaging the optical coating or scratching exposed lens surface, we have de- veloped recommendations for cleaning the lens.
AC power (mains) cord connection Use the supplied power cord to connect your projector to the wall outlet. Plug the female power connector into the male connector at the front of the projector. The power input is auto-ranging from 90 to 240 VAC.
4.2.4 Composite Video Input Input specifications The Composite video input section is made of 1BNC input terminal. 1.0 Vpp ± 3dB 75 Ω terminated No loop through 1. data or video R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 4. Connections...
Standard S-Video (S-VHS) with separate Y(luma) and C (chroma) signals. How to select the S-Video input ? 1. Press 4 on the RCU Note: Another way for selecting this input is via Source on the local keypad or via the Menu. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
DVI-I type connector analog link (4 pins at the right side of the connector) not sup- ported Image 4-8 How to select the DVI Input ? 1. Press 5 on the RCU Note: Another way for selecting this input is via the Menu. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 4. Connections...
4.3.1 RS232 IN connection What can be connected to the RS232 IN connection ? The RS 232 IN connections allows the projector to communicate with a Computer e.g. IBM PC or Apple Macintosh. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
• wide range of control possibilities Data communication: sending data to the projector or copying the data from the projector to the computer To set up the baudrate of the projector, see the Installation menu 4.4 Connecting the MGP in Dicom Theatre...
In a dual head setup, the two projector are default programmed with respectively projector address 1 and 2. In order to control the right projector , the executive RCU has to be programmed with the right address. This done by pressing the numeric key 1 or 2 depending on the projector to be controlled.
IMAGE M G P A R CO Image 5-3 If the projector doesn’t go in operational mode verify that the main power switch is in the ON position (see Connections) Select a source 1. Push the source button (image 5-4) Each key press moves to the next source.
5.2 RCU & Local keypad How controlling the projector ? The projector can be controlled by the local keypad or by the remote control unit. Location of the local keypad ? The local keypad is located on the topside of the projector.
Remote control functions. This remote control includes a battery powered infrared (IR) transmitter that allows the user to control the projector remotely. This remote control is used for source selection, control, adaptation and set up. It includes automatic storing of picture controls (Bright- ness, Sharpness...) and settings.
’PAUSE’. The image disappears but full power is retained for immediate restarting. standby button, to start projector when the power switch is switched on and to switch off the projector without switching off the power switch.
RCU: R763794K 5.4 Switching on How to switch on. 1. Press the power switch to switch on the projector. When ’0’ is pushed in, the projector is switched off. When ’1’ is pushed in, the projector is switched on The projector starts in standby mode, LED3 is red.
The total lifetime of the lamp for a safe operation is 1500 hours max, do not use it longer. Always replace with a same type of lamp. Call a BARCO authorized service technician for lamp replacement. When the lamp runtime reaches 1500 hours the projector switches automatically to the other lamp, being lamp2.. following mes- sages are displayed during and after switching.
1. Plug one end of the remote cable in the connector on the bottom of the RCU. 2. Plug the other end in the connector in the front panel of the projector labelled RC. (image 5-14) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Projector address Address installed in the projector to be individually controlled. Common address Default address. Projector will always execute the command coming from a RCU programmed with that common address. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 5. Getting started...
IBM PC (or compatible), Apple MAC, etc. for addresses between 0 and 255. A projector will respond to a RCU set to the common address ’0’ regardless of what address is set in the projector itself (common address of projector should also be “0” ).
PC like menustructure The projector has a build in "PC like" toolbar menu which allows easy access to different parameters for setting up the projector. The menu is activated by pressing MENU, it contains 2 levels depending on the type of user: •...
2. Press ←or → to select the digit to be changed 3. Press ↓ or ↑ to increase or decrease the value To confirm the changes always press ENTER. Use ↓ or ↑ to browse between the different fields. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 43
In some cases an alphanumeric value (file name, ...) has to be entered. Use ↑ or ↓ to scroll through the char- acter values once the input field is activated Following characters can be browsed in this particular order: Decimal scroll list: 0123456789 Signed decimal scroll list: 0123456789- ASCII scrolllist:ABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789+-*/&@#.;.abcdefghijklmnopqrstuvwxyz R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 44
5. Getting started R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
In medical imaging, it is important that there be a visual consistency in how a given image appears, whether viewed, for example, on the display monitor of a workstation, a projector or as a film on a light-box. In the absence of any...
Analogy to the typical architecture • Display function : the function can be chosen to be DICOM, DIN, standard or custom. This is done in the projector. • Preset: is determined by the user on MediCal •...
Page 47
FACTORY RESET PASSIVE MEMORY Image 6-2 It is recommended to use the DICOM presets since they give the best grayscale for medical applications. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 6. Introduction to Medical Visualization systems Color temperature DICOM Clearbase...
Page 48
6. Introduction to Medical Visualization systems R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
4. Use ↑ or ↓ to select Autosetup 5. Press → to pull down the menu (image 7-1) 6. Use ↓ or ↑ to select the desired method 7. Press ENTER Image 7-1 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 7. Main Menu...
Page 50
7. Main Menu R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
4. Use ↑ or ↓ to select Contrast (image 8-1) 5. Press ENTER A sliderbox is displayed 6. Use ←or →, the numeric keys on the remote, or the keypad to change the Contrast. (image 8-2) Image 8-1 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 8. Adjustments...
6. Use ←or →, the numeric keys on the remote, or the keypad to change the Brightness. Image 8-3 Brightness adjustment is only possible if the Display function is Standard. In the other modes (DICOM, DIN, CUSTOM ) the brightness adjustment is disabled. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
A sliderbox for the Red custom setting is displayed as well as a wizard textbox in the lower part of the screen. (image 8-5, image 8-6) 7. Use ←or →, the numeric keys on the remote, or the keypad to change the Color temperature. 8. Follow the instructions on the wizard textbox. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 8. Adjustments...
The “show native resolution” function will handle the sources as shown in the table. • Full screen: The Full screen method forces to use the complete native resolution of the LCD panels independently of the native resolution of the source. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
7. Press ENTER Image 8-7 8.7 Aspect ratio What can be done ? The aspect ratio setting forces the projector to project an image using a defined aspect ratio. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 Projected image Ratio Resolution...
How to reposition the image ? 1. Press MENU to activate the Toolbar 2. Press → to select the Adjustments item 3. Press ↓ to Pull down the menu 4. Use ↑ or ↓ to select Positioning R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
6. Use ↑ or ↓ to select colour (image 8-10) 7. Press ENTER A sliderbox is displayed 8. Use ←or →, the numeric keys on the remote, or the keypad to change the colour. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 8. Adjustments...
Page 58
6. Use ↑ or ↓ to select Tint (image 8-11) 7. Press ENTER A sliderbox is displayed 8. Use ←or →, the numeric keys on the remote, or the keypad to change the Tint. Image 8-11 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 59
6. Use ↑ or ↓ to select Noise Reduction (image 8-13) 7. Press ENTER A sliderbox is displayed 8. Use ←or →, the numeric keys on the remote, or the keypad to change the Noise Reduction. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 8. Adjustments...
Input filtering The input section of the projector allows an additional filtering of the RGB signal by means of a 4 level (0–4) filter setup and can result in a more stable and crisp image, especially with respect to phase stability.
Image 8-14 8.11 Standby Timer Purpose of the Standby Timer If there is no signal, and the standby timer is enabled, a dialogbox is displayed and the projector will shut down after a determined time. Image 8-15 The countdown time can be set in a dialog box in a range from 180 to 3600 seconds (default value = 300). The Timer can also be disabled.
Disabled Countdown to Image 8-17 8.12 Blanking What can be done ? The image can be blanked in several ways : • Top blanking • Bottom blanking • Left blanking • Right blanking R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 63
4. Use ↑ or ↓ to select Blanking (image 8-19) 5. Press ENTER A sliderbox is displayed (image 8-20) 6. Use ←or →, the numeric keys on the remote, or the keypad to change the blanking. Image 8-19 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
6. Use ↑ or ↓ to select the desired parameter, a box surrounds the selected item, press ENTER to activate 7. Press MENU or BACK to exit or to go back to the previous menu R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 65
Image 8-21 Autosetup parameters Frequency Phase Geometry Gain Image 8-22 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 8. Adjustments...
: icon representing the shortcut numeric key to be entered on the RCU to select the same input • a Barco icon : indicates the presence of that particular source. Another way of selecting an input is via the RCU using the numeric keys or via the local keypad.
9.2 Displaying the Logo Logo The projector is factory fitted with a standard Barco Logo (SXGA resolution), however it can be replaced by a new corporate logo using the Take screenshot function. Logo display is done automatically in case of signal absence under the condition that Background is set to Logo A standby timer can also be set to put the projector in standby after a predetermined time in case of signal absence.
9.3 Pause Interrupting the image projection With the Pause function, the image projection can be stopped, the projector remains with full power for immediate restart. The projection is interrupted by means of a mechanical shutter cutting the light beam. How to interrupt the image projection ? 1.
4. Use ↑ or ↓ to select Freeze (image 9-5) 5. Press ENTER to activate the Freeze function Image 9-5 The image can also be frozen using the FREEZE key on the RCU R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
CIE graph coordinates: x = 0.25 ; y = 0.305 Since the projector is operating most of the time in greyscale mode meaning a B/W signal is to be reproduced, the color temperature setting can be seen as an adjustment of the neutral greyscale color temperature.
10.1.3 Color temperature in a closed loop medical visualization system Color temperature in a closed loop medical visualization system If, however the projector is part of a typical closed loop medical visualization system (Barco Dicom will be set by the control & calibration system (MediCal The preset (formed by color temperature and display function) values are checked on a regular basis to verify if the projected image is within a tolerance with respect to consistency, in the case it isn’t, a calibration process will be launched.
In a standalone or open loop system i.e. without any feedback from the screen, the display function is defined by the user via the OSD of the projector. The selected display function corresponds then to a Look Up Table (LUT) as it is was set in the factory.
10.2.2 Display function in a closed loop medical visualization system Display function in a closed loop medical visualization system If, however the projector is part of a typical closed loop medical visualization system (Barco Dicom will be set by the control & calibration system (MediCal How to select a Display function ? 1.
& white BRG to Y black & white BGR to Y black & white R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 Comment Green is replaced by white (bright) Red is replaced by white (bright) Blue is replaced by white (bright)
Page 76
5. Press → to pull down the menu 6. Use ↑ or ↓ to select the desired selection 7. Press ENTER Image 10-9 RGB to Y Image 10-11 G to Y Image 10-13 RBG to Y R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
The User control parameter allows to lock the RCU, the commands coming from the RCU (or the executive RCU) are then ignored in the projector. This can be used to prevent unauthorized adjustment of projector parameters. How to lock the RCU ? 1.
Single mode The projector will always switch to the lamp with the shortest runtime when the difference between the runtimes of lamp1 and lamp 2 reaches 100 hours, switching from one lamp to another happens only at switching on of the projector and not during operation.
6. Use ↑ or ↓ to select Runtime warning (image 10-18) 7. Press ENTER A dialogbox is displayed (image 10-19) 8. Use ←or →, the numeric keys on the remote, or the keypad to change the runtime warning setting. Image 10-18 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 10. Settings...
10. Settings Image 10-19 ARNING Lamp runtime reset as well as the lamp replacement can only be done by a Barco authorized technician. 10.8 Input Balance AUTION The input balance adjustmentis only for Barco authorized technicians ! Introduction: Unbalanced color signals When transporting signals, there is always a risk of deterioration of the information contained in the signals.
Page 81
100 % black (background) full screen pattern Image 10-23 White balance : In the projector, we will set the contrast for each color until we get a 100% light output picture when projecting a 100% white image (image A) Black balance : In the projector, we will set the brightness for each color until we get a 0% light output picture when projecting a 100% black image (image B).
12.Adjust the Blue black level until bright spots appear on the screen 13.Adjust the red black level until bright spots appear on the screen The projected image should know be noisy full black Image 10-25 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 83
12.Adjust the Blue white level (gain) until bright spots appear on the screen 13.Adjust the Red white level (gain) until bright spots appear on the screen the projected image should know be noisy neutral grey. Image 10-28 Image 10-29 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 10. Settings...
Beside the AGC there is the possibility to manually set the gain of the incoming video signal. When the AGC is enabled (ON), the manual setting does not affect the gain, AGC must therefore be disabled. The manual gain control must be done on an external pattern with white areas (grey scale bar pattern) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
4. Use ↑ or ↓ to select Lamps 5. Press → to open the menu (image 10-33) 6. Use ↑ or ↓ to select History 7. Press ENTER A textbox is displayed (image 10-34) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 10. Settings...
A dialogbox is displayed (image 10-36) 10.Use ←or →, the numeric keys on the remote, or the keypad to change the serial number of the lamp (serial number 0000000 will not be accepted). R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 87
Image 10-35 Image 10-36 ARNING Lamp runtime reset as well as the lamp replacement can only be done by a Barco authorized technician. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 10. Settings...
The Seamless switching mode is not used in the Random mode. Note on Fade in/out In some cases, depending on the sources to be switched, the fade in/out effect is impossible or may not occur. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 11. Tools...
What are the different possibilities within the PiP mode ? The input section of the projector allows a multitude of combinations of different input signals which may be projected in the 4 windows of the PiP screen.
4. Use ↑ or ↓ to select Picture in Picture 5. Press → to pull down the menu 6. Use ↑ or ↓ to select the PiP Select 2. factory layout R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 11. Tools...
4. Use ↑ or ↓ to select Picture in Picture 5. Press → to pull down the menu 6. Use ↑ or ↓ to select the PiP Select 7. Press → to pull down the menu R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
6. Use ↑ or ↓ to select PiP add window (image 11-8) 7. Press ENTER The source selection menu is displayed (image 11-9) In the lower part of the screen appears a wizard in 4 steps (image 11-10) Image 11-8 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 11. Tools...
7. Press ENTER The source selection menu is displayed (image 11-12) In the lower part of the screen appears a wizard in 4 steps (image 11-13) Image 11-11 Image 11-12 Image 11-10 Image 11-13 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Use ← or →, the numeric keys on the remote, or the keypad to enter the name and exit with BACK or MENU. If save as has been selected, a messagebox is displayed (image 11-16) Image 11-14 Image 11-15 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 Image 11-16 11. Tools...
The fixed layouts and the active layout can not be deleted. How to delete a layout ? 1. Press MENU to activate the Toolbar 2. Press → to select the Tools item 3. Press ↓ to Pull down the Tools menu Image 11-18 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
PiP adjust in case a particular 4 window layout Data1 window has the focus, new settings will only affect Data1 window Video1 window has the focus, new settings will only affect Video1 window R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 11. Tools...
Follow the procedure. 11.8 Background Purpose If there is no signal connected to the projector, the background will be a logo, a black or a blue screen depending on the background settings. How to change the background ? 1. Press MENU to activate the Toolbar 2.
The logo display is only possible in the full screen mode, a logo can thus not be displayed (rescaled) in a window in the PiP mode. When there is no signal connected, the projector will also start its standby timer countdown (if enabled) and shuts down after the predetermined time.
Introduction to video formats An image file contains the main characteristics of a source (number of active lines,...). The projector’s memory contains a list of files corresponding to the most common sources, these are the standard files (file extension= *.s). When a new source corresponds to one of these files, a custom file (file extension=*.C) is created and saved for future use.
A dialogbox is displayed (image 12-5) 9. Use ←or →, the numeric keys on the remote, or the keypad to edit and change the values, confirm with ENTER Note: greyed out fields can not be updated (total pixels) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Vertical start in lines Advanced video settings The advanced button enables the advanced settings for a video source. Image 12-6 The Comb filter is by default enabled. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 Image 12-5 video signal frame blanking egalisation...
4. Use ↑ or ↓ to select Rename ® disturbances since the AGC is hold during a long interval, thus reducing the ® pulse. video info Image 12-9 Hs horizontal sync pulse active low active high R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
A textbox is displayed (image 12-14) Use ←or →, ↓ or ↑ on the remote, or the keypad to enter the new name, confirm with ENTER. Image 12-12 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 Image 12-11 12. Video formats...
A dialogbox is displayed (image 12-16) 6. Use ↑ or ↓ to select the desired file 7. Press ENTER The selected file is deleted and is removed from the list Image 12-15 Image 12-14 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 107
12. Video formats Image 12-16 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 108
12. Video formats R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
2. Press → to select Information 3. Press ↓ to Pull down the menu 4. Use ↑ or ↓ to select Configuration (image 13-3) 5. Press ENTER A textbox appears on the screen. (image 13-4) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 13. Information...
2. Press → to select Information 3. Press ↓ to Pull down the menu 4. Use ↑ or ↓ to select Current input format (image 13-6) 5. Press ENTER A textbox appears on the screen. (image 13-7) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
3. Press ↓ to Pull down the menu 4. Use ↑ or ↓ to select Diagnostics 5. Press → to select I2C (image 13-8) 6. Press ENTER A textbox appears on the screen. (image 13-9) Image 13-8 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 13. Information...
2. Press → to select the Information 3. Press ↓ to Pull down the Information menu 4. Use ↑ or ↓ to select Quick access keys (image 13-11) 5. Press ENTER A textbox appears on the screen. Slave address R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Page 113
2. Push the desired function key for 3 seconds The menu item is stored in the quick access key Factory pre-programmed functions -F1 : color depth -F2 : noise reduction -F3 : orientation R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 13. Information...
Page 114
13. Information R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
5. Press → to open the menu 6. Use ↑ or ↓ to select Zoom/Focus or Shift (image 14-1) 7. Press ENTER A textbox appears on the screen, follow the instructions. (image 14-2, image 14-3) R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 14. Installation...
What can be done ? In some cases the projector address must be changed, for example if an unique RCU is used to control 2 or more projectors. The projector address setting on the projector must match the setting on the RCU.
Image 14-5 14.3 Orientation Projector orientations Depending on how the projector is oriented, the projector’s internal settings have to be adapted. How to change the orientation ? 1. Press MENU to activate the Toolbar 2. Press → to select the Installation item 3.
The view mode of the MGP 10 can be adapted to be Portrait or Landscape. Both view mode are made possible thanks to the corresponding ceiling mount. Portrait/Landscape setting is done once upon installation, see "Projector configurations", page 12 How to change the view mode ? 1.
Image 14-7 14.5 Language List of languages The list of selectable languages is depending on the software verison of the projector. How to change the Language? 1. Press MENU to activate the Toolbar 2. Press → to select the Installation item 3.
A security function is implemented in the projector and allows a protection against theft. A PIN code allows the user to lock the projector in case of wrong code entry. The PIN code must be entered at each start up (Power ON), entering three times a wrong number triggers a wait cycle of 15 minutes, the second 3 wrong codes a wait cycle of 30 minutes, 1 hour, ...
How to disable the security mode ? 1. Press MENU to activate the Toolbar 2. Press → to select the installation item 3. Press ↓ to Pull down the menu R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 Image 14-12 Image 14-14 14. Installation...
A dialogbox is displayed. (image 14-17) 6. Use ← or → , the numeric keys on the remote , or the keypad to enter and confirm the new password. Each character is displayed as an asterisk. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
Image 14-16 Image 14-17 The new password is accepted if the new password and the confirmed password coincide. If the password is forgotten contact a Barco authorized technician. 14.9 Auto Menu Exit Purpose The Auto menu exit allows to quit automatically the menu after 1 minute of inactivity.
14.10Automatic startup What can be done ? The automatic startup allows immediate restart of the projector after a power failure (breakdown), i.e. without passing through the standby state. The projector restarts at power resume and recovers the previous settings (previous source,...).
4. Use ↑ or ↓ to select Textbox 5. Press → to pull down the menu 6. Use ↓ or ↑ to enable/disable the textbox (image 14-20) 7. Press ENTER Image 14-20 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 14. Installation...
3. Press ↓ to Pull down the Installation menu 4. Use ↑ or ↓ to select Status bar position (image 14-22) 5. Press ENTER 6. Use ↑ or ↓ to position the status bar R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
4. Use ↑ or ↓ to select Sliderbox position (image 14-23) 5. Press ENTER A sliderbox is displayed. Use the 4 arrow keys to drag the box to the desired position. (image 14-24) Image 14-23 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004 14. Installation...
Page 128
14. Installation Image 14-24 There is a coarse and a fine adjustment of the position, use ENTER (when sliderbox is displayed) to switch between the two. R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004...
15.2 Cleaning How to clean the dustfilters ? 1. Place the projector so as to access easily the filter to be cleaned 2. Push the handle downwards to unlock the filter. (image 15-2) 3. Slide out the filter (image 15-3, image 15-4, image 15-5) See image 15-6.
Page 130
At that time it is strongly recommended to replace or clean the dustfilter under the X-cube. Failing to do so, will strongly reduce the lifetime of the LCD’s and the analyzers. The manufacturer reserves itself the right to refuse warranty repair if the projector was working with dirty dustfilters.
16. STANDARD IMAGE FILES 16.1 Table overview Table overview The following standard image files are pre-programmed in the projector. Name Resolution 1600_48V 1600x600i 640x200i 1024x480i COMPUSC4 DOS1_70 720x400 DOS3_56 640x400 DOS4_85 640x400 640x350 ESVGA_75 832x624 EXGA_60 1152x864 1152x864 EXGA_80 EXGA_85...