Main Specification - FujiFilm PB2020-M Operation Manual

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Main Specification

<Classification of Medical Electrical Equipment>
1. Type of protection against electric shock
2. Degree of protection against electric shock
3. Degree of explosion protection
4. Degree of protection against ingress of water : IPX7
[Note] Use in combination with the SP-900 ultrasonic processor.
<Applied Part>
Insertion portion
<Specifications>
202B1263042
Working length
Outside diameter of insertion
portion
Maximum diameter of
insertion portion
Axial resolution
Lateral resolution
Penetration depth
Acoustic frequency
Spatial-peak temporal-
average intensity
Mechanical Index (MI)
: Class I equipment
(power supply: protective earth plug)
: Type BF applied part
: Use is prohibited in an oxygen-rich
environment or in a flammable gas
atmosphere.
2150 mm
1.4 mm to 1.9 mm
2.0 mm
2 mm or less
2 mm or less
7 mm or more
(assuming the attenuation coefficient of 0.3 dB cm
20 MHz ±15%
Izpta. α ≤ 720 mW cm
Less than 1.0
Main Specification
-1
MHz
-2
-1
-1
(α=0.3 dB cm
MHz
)
-1
)
53