FujiFilm PB2020-M Operation Manual page 28

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Chapter 3 Inspection Before Use
(4) Press the [FREEZE] key on the keyboard of the
ultrasonic processor to disable freeze mode.
Make sure that the transducer of the ultrasound
head is rotating and that a multiple-echo image
appears on the monitor.
[Note]
Make also sure that the insertion portion and
other components function normally and do not
produce abnormal sound.
(5) Make sure that the image on the monitor
shakes when shaking the endoscopic ultrasonic
probe's distal end in sterile water.
(6) Press the [FREEZE] key on the keyboard of the
ultrasonic processor to enable freeze mode.
[Note]
Make sure that the transducer of the ultrasound
head stops rotating.
28
202B1263042