FujiFilm PB2020-M Operation Manual page 57

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Electromagnetic immunity compliance information and guidance
EN 60601-1-2
Immunity test
Conducted RF
3Vrms
EN 61000-4-6
150kHz to 80MHz
Radiated RF
3V/m
EN 61000-4-3
80MHz to 2.5GHz
Electromagnetic immunity compliance information and guidance
The customer or the user of this product can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitter) and this product as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter P (W)
0.01
0.1
1
10
100
202B1263042
Compliance
Test level
level
3V[V
]
1
3V/m[E
]
1
Separation distance related to frequency of transmitter (m)
150kHz to 80MHz
d=1.2
√ P
0.12
0.38
1.2
3.8
12
Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of this product,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of transmitter.
Recommended separation distance
3.5
√ P
d=
V
1
3.5
80 to 800MHz
d=
√ P
E
1
7
800MHz to 2.5GHz
d=
√ P
E
1
Where "P" is the maximum output power rating of
transmitter in watts (W) according to the transmitter
manufacturer and "d" is the recommended
separation distance in meters (m).
This product complies with the requirements of
EN 60601-1-2: 2007.
However electromagnetic interference may occur on
this product under electromagnetic environment that
exceeds its noise level.
Electromagnetic interference may occur in the
vicinity of equipment marked with the following
symbol.
80 to 800MHz
d=1.2
√ P
0.12
0.38
1.2
3.8
12
Main Specification
800MHz to 2.5GHz
d=2.3
√ P
0.23
0.73
2.3
7.3
23
57

Advertisement

Table of Contents
loading

Table of Contents