FujiFilm PB2020-M Operation Manual page 31

Endoscopic ultrasonic probe
Table of Contents

Advertisement

When performing an ultrasound examination with this product, it is necessary to
have the ultrasound head directly contact the region of interest or have an ultrasound
transmission medium such as sterile water between the ultrasound head and the
region of interest.
(1) W h e n n e c e s s a r y , i n j e c t a n u l t r a s o u n d
transmission medium such as sterile water to
the region of interest, for which an ultrasound
examination is performed.
[Note]
Injecting an excessive amount of an ultrasound
transmission medium such as sterile water may
cause accidental swallowing or vomiting.
(2) Unlock the bending section by turning the
up-down and left-right angulation locks.
(3) Operate the up/down and left/right angulation
knobs to straighten the bending section to its
neutral position.
202B1263042
Do not push or pull this product or insert into the bending portion of an
endoscope when this product is operating (freeze mode is disabled). If the
endoscope is bent excessively or the forceps stand is moved up, slowly
push or pull this product. Pushing or pulling the endoscope forcibly or
suddenly may cause distortion of the image or damage to this product.
It is desirable that the endoscope is straight and the forceps stand is in
neutral position before operating this product. Operating this product
under unfavorable conditions may cause distortion of the image or
damage to this product.
CAUTION
Chapter 4 Method of Use