FujiFilm PB2020-M Operation Manual page 14

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Chapter 1 Precautions
CAUTION
Clean and disinfect (or sterilize) this product in a clean environment.
If dust adheres to this product after cleaning and disinfection (or
sterilization), it may enter the patient's body.
After cleaning, rinse off any remaining detergent solution with water. If
detergent solution remains, chemicals to be used in the subsequent steps
may be diluted or adulterated.
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, it may flow into the patient's body.
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, the chemical disinfection to be performed at
the next step will lose its effectiveness.
After immersing in chemical solution, rinse off the remaining disinfectant
solution with sterile water. If disinfectant solution remains, it may flow into
the patient's body.
Proceed with aeration after gas sterilization. If this product is used without
performing aeration, gas remaining in this product after gas sterilization
may harm human body.