Chapter 5 Cleaning
(5) Clean the external surface (insertion portion,
b o o t , p r o b e c o n n e c t o r , e n d h o l d e r a n d
connector cap) of the endoscopic ultrasonic
probe with clean gauze in detergent solution.
(6) L e a v e t h e e n d o s c o p i c u l t r a s o n i c p r o b e
i m m e r s e d i n t h e d e t e r g e n t s o l u t i o n i n
accordance with temperature, concentration
and time recommended by the detergent
solution's manufacturer.
(7) Remove the endoscopic ultrasonic probe from
detergent solution.
(8) Wipe any detergent solution remaining on the
external surface of the endoscopic ultrasonic
probe with clean lint-free gauze.
(9) If the endoscopic ultrasonic probe is still dirty,
repeat steps (1) to (8).