6.3 Chemical Disinfection of Endoscopic Ultrasonic Probe

(1) A t t a c h t h e c o n n e c t o r c a p t o t h e p r o b e
connector.
(2) Fill a clean basin with a lid with disinfectant
solution at the temperature and concentration
recommended by the disinfectant solution's
manufacturer.
(3) Fully immerse the endoscopic ultrasonic probe
in disinfectant solution.
[Note]
P e r f o r m a l l d i s i n f e c t i n g s t e p s w h i l e t h e
endoscopic ultrasonic probe is fully immersed
in disinfectant solution.
(4) Wipe the external surface (insertion portion,
b o o t , p r o b e c o n n e c t o r , e n d h o l d e r a n d
connector cap) of the endoscopic ultrasonic
probe with sterile gauze in disinfectant solution
to remove air bubbles completely.
(5) Wipe the probe connector with sterile gauze
saturated with disinfectant solution 3 times
w h i l e r o t a t i n g b a c k a n d f o r t h t h e p r o b e
connector.
[Note]
Ensure the connector cap is securely attached
and is not loosened when wiping the probe
connector.
202B1263042
Chapter 6 Chemical Disinfection
45