Chapter 8 Storage
52
Do not store this product in a packaging box. Doing so may cause
infection.
Do not store this product in locations that do not satisfy storage
conditions. Otherwise, a failure may result.
Store the endoscopic ultrasonic probe that has been disinfected in a location that
meets the following conditions.
Storing conditions
Temperature
: -20 to +60°C
Humidity
: 10 to 85%RH (no dew condensation)
Pressure
: 70 to 106 kPa (within range of atmospheric pressure)
State of the endoscopic ultrasonic probe :
Hang this product with its distal end down.
Keep out of direct sunlight, UV light or X-ray.
[Note] Keep more than 250 mm diameter to store it round.