Inspecting For Combination - FujiFilm PB2020-M Operation Manual

Endoscopic ultrasonic probe
Table of Contents

Advertisement

3.3 Inspecting for Combination

(1) Align the marks on the probe connector and the
scanner, and insert the scanner until it clicks.
[Note]
To remove the probe connector, pull it straight
out.
[Note]
Make sure that the probe connector is not wet.
The wet probe connector may cause equipment
failure.
(2) Insert the insertion portion of this product
through the end holder so as to keep the
ultrasound head and insertion portion from
touching the floor.
[Note]
In clinical procedures in which the endoscopic
ultrasonic probe that has been sterilized is
used, do not use the end holder and place
it temporarily using a method that suits the
clinical procedure.
(3) Make sure that the ultrasonic frequency of
the connected endoscopic ultrasonic probe is
displayed in the "FREQ" area on the right side
of the monitor screen.
[Note]
If the probe cord is dirty, the frequency may not
be accurately displayed.
202B1263042
Do not detach the endoscopic ultrasonic probe from the prove drive unit
while freeze mode is disabled (the transducer is rotating). Doing so may
cause equipment failure of the endoscopic ultrasonic probe or the prove
drive unit.
Chapter 3 Inspection Before Use
CAUTION
27

Advertisement

Table of Contents
loading

Table of Contents