Chapter 1 Precautions
WARNING
Be sure to wear sterile gloves when inspecting this product. Not doing so
may contaminate this product and pose an infection risk.
In clinical procedures in which this product that has been sterilized is
used, exercise caution not to contaminate this product when placing it
temporarily. Not doing so may pose an infection risk.
Slowly insert this product straight into the endoscope. Also, when
withdrawing it, slowly pull straight out. If it is inserted or withdrawn quickly,
body fluid may be splattered around due to breakage or accidental
detachment of the instrument, leading to infection.
Slowly pull this product out of the end holder. If it is withdrawn quickly
from the end holder, body fluid may be splattered around, posing an
infection risk.
Be sure to clean and disinfect (or sterilize) this product as per instructions
as per instructions in this manual after every case. If this product is
not disinfected (or sterilized), it may pose an infection risk or cause an
accident.
If the image often becomes dark during observation, air bubbles may be
on the transducer inside the ultrasound head. However, do not remove air
bubbles once this product is inserted. Body fluid adhered to this product
may be splattered, posing an infection risk.
Do not withdraw this product sharply. In so doing, body fluids adhered to
this product may be splattered, posing an infection risk.
This product is not compatible with cleaning and disinfection with
an automated endoscope reprocessor. If an automated endoscope
reprocessor is used, effective cleaning and disinfection cannot be
achieved, and it may pose an infection risk.
Excessive foaming prevents detergent solution from sufficiently contacting
the surfaces and channel walls of the endoscope, and may impair
effective cleaning.
Do not reuse clean water and detergent solution that were used for
cleaning. If reused, effective cleaning cannot be achieved, and it may
pose an infection risk.