Chapter 5 Cleaning; Method Of Cleaning - FujiFilm PB2020-M Operation Manual

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Chapter 5 Cleaning

This chapter explains how to clean this product.
Pre-cleaning is performed at the bedside immediately after an examination is
completed.
Manual cleaning is performed in a sink, etc. after pre-cleaning.

5.1 Method of Cleaning

Use potable water or detergent solutions to clean this product. For detergent solutions,
refer to "5.2 Detergent Solution".
Table 5.1 Applicable methods of cleaning for the endoscopic ultrasonic probe
202B1263042
This product is not compatible with cleaning and disinfection with
an automated endoscope reprocessor. If an automated endoscope
reprocessor is used, effective cleaning and disinfection cannot be
achieved, and it may pose an infection risk.
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, it might flow into the patient's body.
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, the chemical disinfection to be performed at
the next step will lose its effectiveness.
Chemical cleaning
Yes
WARNING
CAUTION
Ultrasonic cleaning
No
Chapter 5 Cleaning