Chapter 7 Sterilization; Methods Of Sterilization - FujiFilm PB2020-M Operation Manual

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Chapter 7 Sterilization

This chapter explains how to sterilize this product.
Perform ethylene oxide gas sterilization for this product.
This product can withstand up to 30 times of ethylene oxide gas sterilization.
Ethylene oxide gas sterilization applies more load to this product than chemical
cleaning and disinfection, and reduces the product life.

7.1 Methods of Sterilization

To enhance the effect of sterilization, perform cleaning before gas sterilization
according to the procedures described in this manual.
Table 7.1 Applicable methods of sterilization for the endoscopic ultrasonic probe
202B1263042
Proceed with gas sterilization after vaporizing water out of this product.
If gas sterilization is performed while moisture remains in this product,
sterilization of wet parts will be incomplete.
Proceed with aeration after gas sterilization. If this product is used without
performing aeration, gas remaining in this product after gas sterilization
may harm human body.
Do not use boiling or autoclaving for this product. Otherwise, this product
will be damaged by high temperature.
Autoclaving
No
WARNING
CAUTION
CAUTION
Ethylene oxide gas
sterilization
Yes