Chapter 1 Precautions

1.3.3 Direct Harm to Human Body

WARNING
To avoid the potential for patient injury including perforation and bleeding,
do not apply excessive force of this product against walls of the bronchus
and digestive tract. Only advance this product under direct visualization.
When using this product for an endoscope with two instrumaent channels,
do not use an electrosurgical unit at the same time. Doing so may cause
thermal injury to a patient and/or end-user.
Do not hold or scratch this product with metallic forceps, needle, basket,
etc. Doing so may damage the ultrasound head and result in damage to
tissues inside the body cavities or leakage of the ultrasound transmission
medium to the body cavities.
Do not insert this product into an endoscope when a clear endoscopic
image is not obtained on the monitor or when an endoscopic image is
frozen. Doing so may cause damage to tissues in the body cavities,
bleeding, perforation or damage to the equipment.
When using this product under fluoroscopy, note that a contrast image
of the region within approx. 5 mm from the distal end of the endoscopic
ultrasonic probe is not obtained. If this product is inserted forcibly, it may
cause bleeding or perforation.
Slowly insert this product into the endoscope. Inserting it quickly may
cause damage to tissues in the body cavity, bleeding or perforation.
This product must be properly aerated after ethylene oxide gas
sterilization to remove toxic components. Gas remaining in this product
after ethylene oxide gas sterilization may harm human body.