FujiFilm PB2020-M Operation Manual page 56

Endoscopic ultrasonic probe
Table of Contents

Advertisement

Main Specification
Electromagnetic immunity compliance information and guidance
Immunity test
Electrostatic discharge
(ESD)
EN 61000-4-2
Electrical fast transient/
burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11
Power frequency (50/60
Hz) magnetic field
EN 61000-4-8
56
EN 60601-1-2
Test level
± 6kV: contact
± 8kV: air
± 2kV: for power supply
lines
± 1kV: for input/output lines
± 1kV: Line to line
± 2kV: Line to ground
< 11.5V (> 218.5V dip)
For 0.5 cycle
92V (138V dip)
For 5 cycle
161V (69V dip)
For 25 cycle
< 11.5V (> 218.5V dip)
For 5 sec
3 A/m
Compliance
level
Floors should be wood, concrete,
or ceramic tile. If floors are
Same as
covered with synthetic material,
left
the relative humidity should be at
least 30%.
Main power quality should be
Same as
that of a typical commercial or
left
hospital.
Main power quality should be
Same as
that of a typical commercial or
left
hospital.
Main power quality should be
that of a
typical commercial or hospital. If
the user of this product requires
Same as
continued
left
operation during power mains
interruptions, it is recommended
that this product is powered from
an uninterruptible power supply
or battery.
It is recommended to use this
Same as
product by maintaining enough
left
distance from any equipment that
operates with high current.
Guidance
202B1263042