WARN IN G
WARN IN G
WARN IN G
WARN IN G
CAU T ION
CAU T ION
Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
The CSA mark is a registered trademark of the Canadian
Standards Association.
The C-Tick mark is a registered trademark of the
Australian Spectrum Management Agency.
This text denotes the instrument is an Industrial Scientific
ISM1-A
and Medical Group 1 Class A product.
If this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a
normal condition (in which all means for protection are intact)
only.
No operator serviceable parts inside. Refer servicing to qualified
service personnel. To prevent electrical shock do not remove
covers.
This is a Safety Class 1 Product (provided with protective earth).
The mains plug shall only be inserted in a socket outlet provided
with a protective earth contact. Any interruption of the protective
conductor inside or outside of the instrument is likely to make the
instrument dangerous. Intentional interruption is prohibited.
To prevent electrical shock, disconnect the instrument from mains
before cleaning. Use a dry cloth or one slightly dampened with
water to clean the external case parts. Do not attempt to clean
internally.
This product complies with overvoltage Category II and Pollution
Degree 2.
VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be
restricted. The ambient temperature (outside the cabinet) must be
less than the maximum operating temperature of the product by 4 °C
for every 100 watts dissipated in the cabinet. If the total power
dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.