Broadband Component

(Isolator, Circulator, Coupler, Attenuator and so on)
Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
Characteristics
• Wide wavelength measurement range
• Usually low CD and PMD, but especially PMD may need to be
verified
• Loss may be high, but relatively independent of wavelength
Measurement Tips
The broadband nature of these devices allows you to trade off
wavelength resolution for faster measurement speed.
• Set the modulation frequency in manual mode to 2000 MHz.
• In either swept or stepped mode, optimize measurement
speed by setting the wavelength increment just small enough
to resolve the wavelength dependencies of the attributes
being measured.
• If you use the smoothing function to reduce trace noise, you
may need to increase the number of trace points. (for
example, reduce the wavelength increment) to avoid loss of
resolution.
91