Agilent Technologies 86038B User Manual page 337

Photonic dispersion and loss analyzer
Table of Contents

Advertisement

Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
Zero dispersion slope uncertainty
Specifies the uncertainty of the measurement results for
dispersion slope"
, expressed as the maximum difference
between the measured value and actual value.
Conditions: valid for measurements on fibers. Zero dispersion
slope derived from a suitable fit (quadratic or Sellmeier) to the
chromatic dispersion data measured around the zero dispersion
wavelength. Measured wavelength range for fitting as specified.
Wavelength increment as specified. Other conditions: same as
for
"Relative group delay (rGD)"
uncertainty"
.
Measurement: The specification is derived from the
dispersion (CD) uncertainty"
wavelength uncertainty"
Zero-dispersion wavelength
Wavelength where the
under test equals zero.
Zero-dispersion wavelength uncertainty
Specifies the uncertainty of the measurement results for ¨zero-
dispersion wavelength, expressed as the maximum difference
between the measured value and actual value.
Conditions: valid for measurements on fibers. Zero dispersion
wavelength derived from a suitable fit (quadratic or Sellmeier)
to the chromatic dispersion data measured around the zero
dispersion wavelength. Measured wavelength range for fitting as
specified. Wavelength increment as specified. Other conditions:
same as for
"Relative Group delay (rGD) uncertainty"
"Relative wavelength uncertainty"
Measurement: The specification is derived from the
dispersion (CD) uncertainty"
wavelength uncertainty of the used tunable laser source.
and
"Relative wavelength
and the (dynamic)
.
"Chromatic dispersion (CD)"
.
and the (dynamic) absolute
"Zero
"Chromatic
"Relative
of the fiber
and
"Chromatic
337