Agilent Technologies 86038B User Manual page 342

Photonic dispersion and loss analyzer
Table of Contents

Advertisement

Assembled dimensions: (H x W x D)
55.5 cm x 43.5 cm x 55.5 cm
(21.9 in x 17.2 in x 21.9 in)
Power Requirements 100 – 240 V
50 – 60 Hz
~
3 power cables
N5230A: max. 350 W
8164B (incl modules): max. 270 VA
83427C: max. 250 VA
Net weight
Standard system: 54 kg (120 lbs)
Measurement time 30 s
(DGD/PDL swept mode with 4 polarization states, IFBW 70 Hz,
2 nm wavelength range and 10 pm wavelength increment, or 100
nm wavelength range and 1 nm wavelength increment)
Conditions
Maximum Safe Optical Input Power: +4 dBm
Warm-up time: 90 minutes
Storage temperature: -40°C to +70°C
Operating temperature: +10°C to +35°C
Specified temperature: +20°C to +30°C
Humidity: < 80% r.h., non-condensing
Recommended re-calibration period: 1 year
342
Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition