Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
• A second 8164B should have the address 19. A third
8164B should have the address 18, and so on.
• If a special PMF-switch for multi-TLS automation is
included, the unit attached to Port 2 of the switch should
have the address 19, the unit attached to Port 3 should
have the address 18, and so on.
For the Wavelength Meter Option Only
12 Place the 86122A multi-wavelength meter on top of the
PNA.
13 Connect a GPIB cable from the 86122A to the GPIB port
on either the PNA or TLS mainframe.
14 On the 86122A, press Setup > GPIB Setup and set the
address to 10.
15 Connect a patchcord from the 86122A Optical Input to
TLS Output 1.
16 Continue with
"Performance Verification" on page
35.
31