the
deviation trace represents the noise level of
measurement and should be within
0.05 ps.
\
Figure 7
Sample DGD trace for the MXF fiber
b Check the CD curve. The noise level can again be
displayed using Curve Fit and Deviation only settings, as
in Fig. 8. A Quadratic curve should make a good fit, but if
it does not in the measured wavelength range, then a
Sellmeier fit can be used. The deviation represents the
noise level and should be within \ 0.07 ps/nm.
41
Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition