5 Verification and Calibration

Verification Procedure

Introduction

Verification is recommended to verify that the instrument is working properly during incoming
inspection - as an acceptance test. Verification is also recommended if you notice suspicious
instrument behavior or if there is a question about the performance of the instrument. Verification
procedures can be time consuming as they verify the operation of a single channel at a time.
Performing a reference calibration and a channel calibration provides the same high
level of confidence as the manual channel-by-channel verification process.
Verification is not necessary as part of a standard maintenance cycle. It is also not necessary after
performing any of the calibration procedures, as the calibration processes verify that the reference
and channels are functioning within specification.
Verification commands set each channel to a specific output level, which can then be used to verify the
specified accuracy of the channel at that operating point. If the instrument fails any of the verification
tests, try performing the Reference and Channel calibrations again. If verification is still unsuccessful,
return the unit to a Keysight Technologies Service Center.
Verification Equipment and Considerations
A 7½-digit DMM, Keysight 34470A (Run auto-calibration on the DMM prior to starting)
l
BT2152-60002 Verification kit - for BT2152A
l
BT2183A Verification fixture - for BT2152B
l
The ambient temperature is stable and between 20 °C and 28 °C. Relative humidity is <80%
l
The BT2152A/B must be warmed up for at least 1 hour prior to verification tests
l

Verification Fixtures

A multi-channel breakout board assembly can simplify verification when planning to verify all
channels. This can be constructed by wiring all BT2152A/B channels in parallel at the breakout board.
Only one channel at a time can be tested with this arrangement. When selected, that channel is
electrically connected. The remaining channels have internal switches keeping them disconnected
Verification Kit - for BT2152A
The Keysight BT2152-60002 verification kit includes the following parts:
1 - Verification fixture (see below) p/n BT2152-65510
l
8 - Ethernet Cat6A, 26AWG, 8-conductor, 1ft. cables p/n 8121-3032
l
110