Perform a Verification of the U5310A (optional)
Requirements for Verification
The correct operation of the U5310A may be verified by the use of a simple application
which carries out several performance checks on a signal acquired from an external
Function Generator.
Required Hardware
An external signal source is required. Almost any sine wave or function generator
capable of generating a signal with an amplitude of 300 mV rms into 50 Ω at a
frequency of 1 MHz may be used.
Hardware
RF Analog Signal Generator
MMCX cable
Operational Verification Procedure
Do not exceed the maximum voltage level at the INPUT connector (±1.5 Vmax
DC)
1. Configure the RF Generator to produce a Sine signal with a Frequency of 1.0
MHz, an Amplitude of 300 mV rms (+2.55 dBm).
2. Connect the Signal Generator output to the IN 1 connector, and turn on the out-
put.
3. Launch the AgMD2Verify utility from
Windows Start Menu > Keysight MD2 Digitizer > AgMD2Verify.
A command shell window will open.
4. Select the instrument PXI address, then Press any key to start the test. The
screen capture below shows the result of a successful verification:
Keysight U5310A Startup Guide
Description
e.g. Keysight N5181B
Coaxial cable, MMCX (male) to BNC (male), 1 m.
Step 5: Verify Operation of the U5310A Module
21