Current Calibration

Step 1. Put the DMM in current measurement mode on the 10 mA range with a minimum of 100 PLC
averaging. Make sure the DMM's current input terminal is connected to cal port pin 3 as shown in the
previous figure.
Step 2. Start the 10 mA current calibration. Send
Step 3. Trigger a current measurement on the DMM. Measure the current between Com and I.
Cal Step 3
Begins 10 mA current calibration
Step 4. Send the valid measurement to the instrument. Send
If CAL:RES? returns a value other than zero (0), an error has occurred. Send SYST:ERR? to
return the error. Check the calibration setup; then re-run the step.

End Calibration and Store Results

Step 1. End calibration. This returns the instrument to the *RST/Power-on state and sets the
CAL:STEP to 0. Send
CAL:END
Step 2. Store the calibration results. This step overwrites the previous calibration constants. Send
CAL:STOR
Step 3. Disconnect the DMM from the instrument.
Keysight BT2152A/B Operating and Service Guide
CAL:STEP 3
Calibration Limits
0.008 A to 0.011 A
CAL:RES? <value>
5 Verification and Calibration
117