Table of Contents

Advertisement

3 Operating the Analyzer
DIG:PIN1:FUNC DINP
DIG:PIN2:FUNC DINP
DIG:PIN3:FUNC DINP
DIG:PIN4:FUNC DINP
DIG:PIN5:FUNC DINP
DIG:PIN6:FUNC DINP
DIG:PIN7:FUNC DINP  \\ sets the function for all 7 pins to digital input
DIG:PIN1:POL POS
DIG:PIN2:POL POS
DIG:PIN3:POL POS
DIG:PIN4:POL POS
DIG:PIN5:POL POS
DIG:PIN6:POL POS
DIG:PIN7:POL POS
DIG:INP:DATA?

Fault Output

Pins 1 and 2 can be configured as a fault-output pair. The Fault Output function enables a fault
condition to generate a protection fault signal from the
Both pins 1 and 2 are dedicated to this function. Pin 1 is the Fault output; pin 2 is the common for pin 1.
This provides for an optically-isolated output. The polarity of pin 1 can also be configured. When the
pin polarity is positive, a fault condition causes the isolated output to conduct. Note that the Fault
output signal remains latched until the fault condition is removed and the protection circuit is cleared
using the
OUTput:PROTection:CLEar
Pin 2's selected function is ignored. Connect Pin 2 to the ground of the external circuit.
To configure the Fault Output function:
DIG:PIN1:FUNC FAUL
DIG:PIN1:POL POS

Inhibit Input

Pin 3 can be configured as a remote inhibit input. The Inhibit input function lets an external inhibit
signal disconnect all input channels from the cells. The Inhibit input is level triggered. The signal
latency is 5 microseconds. Pin 8 is the common for pin 3. Use the
specify the following Inhibit input modes: Latching, Live, or Off.
DIG:PIN3:FUNC INH
DIG:PIN3:POL POS
OUTP:INH:MODE LIVE
Note that the front panel Error light turns on when any inhibit signal (live or latched) occurs. To clear
the Error light, remove the inhibit signal and send the
40
\\ sets the function for all 7 pins to positive
\\ reads the data on the pins
command.
\\ sets the function for pin 1 to fault output
\\ sets the function for pin 1 to positive
\\ sets the function for pin 3 to inhibit input
\\ sets the polarity for pin 3 to positive
\\ sets the inhibit mode to Live
Alarm Status Group
OUTput:INHibit:MODE
OUTput:PROTection:CLEar
Keysight BT2152A/B Operating and Service Guide
on the digital port.
command to
command.