Instrument Specifications; System Verification Procedure; Performance Tests - Keysight Technologies N5247B Service Manual

2-port and 4-port pna-x microwave network analyzer
Hide thumbs Also See for N5247B:
Table of Contents

Advertisement

Tests and Adjustments
About System Verification and Performance Tests

Instrument Specifications

The analyzer's instrument specifications are described in the "N5247B Data
Sheet and Technical Specifications" in the section titled "Corrected System
Performance", available online at:
http://literature.cdn.keysight.com/litweb/pdf/N5247-90002.pdf.
These specifications apply when the analyzer is used to make either raw or
error-corrected measurements.

System Verification Procedure

The system verification procedure tests the network analyzer measurement
"system", as defined previously, against the system specifications. If
confirmation is successful, the measurement system is capable of making
measurements to the accuracy specified by the graphs of measurement
uncertainty.
The procedure consists of calibrating the analyzer with a calibration kit,
measuring a set of characterized devices, and comparing the resultant
measured data to the data and uncertainty limits supplied with the verification
kit. The device data provided with the verification kit has a traceable path to
NIST. The total measurement uncertainty limits for the performance
verification are the sum of the factory measurement uncertainties and the
uncertainties associated with measuring the same devices on the system being
verified. The difference between the factory-measured data and the
verification-measured data must fall within the total uncertainty limits at all
frequencies for the total system uncertainty test to pass.
Calibration kits are different from verification kits. Calibration kits are used
to determine the systematic errors of a network analyzer measurement
system. Verification kits are used to confirm system specifications and are
not used to generate error correction.

Performance Tests

Performance tests are used to confirm analyzer performance against the
"instrument" specifications. If confirmation is successful, the PNA meets the
instrument specifications.
Performance tests are contained in the analyzer's firmware with Options
S93898A and are described at
An illustrated outline of the performance verification procedure:
— for ANSI/NCSL Z540.3-2006 and ISO/IEC 17025 verification, is shown in
Figure 3-1 on page
— for non-standards verification, is shown in
3-8
"Performance Tests" on page
3-10.
Keysight N5247B Service Guide
3-32.
Figure 3-2 on page
3-11.