SECTION 4
OPERATION
4.6.3 GENERAL SETUP
GENERAL SETUP is used for UPC SETUP, LCD SETUP, UPC STATUS and POWER SOURCE
STATUS.
From any METER display (V/I, AMPS or POWER), press the fn key to enter SETUP MODE.
From the SETUP display, press 3 to enter GENERAL SETUP 1.
GENERAL SETUP 1:
1 UPC SETUP
2 LCD SETUP
3 UPC STATUS
Press 7 to display these additional items:
GENERAL SETUP 2:
1 INITIAL VOLTAGE
2 KEYBOARD LOCK
3 TRANSIENT WFs
The following pages describe each menu item shown above.
4.6.3.1 UPC SETUP
UPC SETUP is used to select meter sense point, enable or disable Continuous Self Calibration (CSC),
set the Programmable Output Impedance (Option), set transition time and voltage and frequency limits.
From any METER display (V/I, AMPS or POWER), press the fn key to enter SETUP MODE.
From the SETUP display, press 3 to enter GENERAL SETUP 1.
From the GENERAL SETUP 1 display, press 1 to enter UPC SETUP.
UPC SETUP: SENSE=INT, CSC=OFF
PROG.Zo - Hit EDIT. TRANSITION TIME=0
FREQ.
VOLTAGE LIMITS MIN=0
SENSE:
CSC:
4 POWER SOURCE STATUS
5 RANGE CONTROL
6 SLEW RATE SETUP
7 More Options
4
5
6
7
LIMITS MIN=45
INTernal or EXTernal. This is the sense point for metering and CSC.
EXTernal uses the remote sense terminals on the rear panel of the Power
Source.
OFF or ON. Continuous Self Calibration. If EXTernal SENSE is selected, CSC
may be used to compensate for output wiring losses and maintain accurate
program output voltage at the sense point which may be the load.
68
MAX=5000
MAX=600