4.6.2.2 COPY WAVEFORM
COPY WAVEFORM is used to copy waveforms (WF) from one waveform number to another.
From any METER display (V/I, AMPS or POWER), press the fn key to enter SETUP MODE.
From the SETUP display, press 2 to enter WAVEFORM SETUP.
From the WAVEFORM SETUP display, press 2 to enter COPY WAVEFORM.
COPY WAVEFORM:
AMPLITUDE SCALE FACTOR
SOURCE WAVEFORM
DESTINATION WAVEFORM
Press EXECUTE to copy a waveform.
This will cause the UPC to stop generating any output while the waveform is being copied.
AMPLITUDE SCALE FACTOR: Sets the amplitude of the destination waveform relative to the source
SOURCE WAVEFORM:
DESTINATION WAVEFORM:
Example: To put a 10% spike on a sine wave: Copy WF 1 to WF 2 at a 90% scale factor, allowing 10%
more voltage to accommodate the spike. Edit WF 2 adding a 10% spike. (see 4.6.2.1)
From the WAVEFORM SETUP display, press 2 to enter COPY WAVEFORM.
COPY WAVEFORM:
AMPLITUDE SCALE FACTOR
SOURCE WAVEFORM
DESTINATION WAVEFORM
Press 90, ENTER, 1, ENTER, 2, to select amplitude scale factor, source and destination waveforms.
COPY WAVEFORM:
AMPLITUDE SCALE FACTOR
SOURCE WAVEFORM
DESTINATION WAVEFORM
Press EXECUTE to copy the waveform.
The UPC to stop generating any output while the waveform is being calculated and copied. Waveform 2
is now a sine wave but at 90% of the amplitude of waveform 1.
NOTE: When a Program is executed, the UPC will attempt to use the specified waveform at the
programmed voltage. E.g. If the programmed voltage is 100 volts, the above 90% sine wave will
produce an output of 100 V, NOT 90 V. See 4.6.2.1 for an example of using a 90% waveform.
=100
=1
=2
waveform.
1-32 may be copied from. 1 is always a sine wave. 2-16 may be
edited. 17-32 are permanently stored and cannot be changed.
2-16 may be copied to and edited.
=100
=1
=2
=90
=1
=2