Manual Mode; Metering Displays - Pacific UPC-32 Operation Manual

Table of Contents

Advertisement

SECTION 4
OPERATION
4.3

MANUAL MODE

This section describes operation in Manual Mode. In this mode you may slew or directly enter voltage or
frequency values. Parameters other than volts and frequency can only be changed by editing and
executing a steady-state Program.
The V/I METER display will indicate MANUAL MODE in the lower left corner of the display.

4.3.1 METERING DISPLAYS

There are three primary metering displays plus a set of optional Harmonic Analysis displays. Pressing
the DISPLAY key will cycle through the metering displays. If a display other than a metering display is
active, pressing the DISPLAY key will abort common edit functions and return to a metering display.
The V/I METER display indicates FREQuency, metering SENSE point (INTernal or EXTernal) and
(MANUAL MODE or PROGRAM #) executing. This display meters true RMS Voltage (line to neutral and
line to line) and true RMS current.
V/I METER:
FREQ=60.00
SENSE=INT
MANUAL MODE Ia=00.00
The POWER METER display meters kVA, true kW and PF.
POWER METER: PHASE A
KVA
KW
PF
The AMPS METER display meters RMS amps, PEAK amps and CREST FACTOR.
AMPS METER: PHASE A
RMS
PEAK
CREST FACTOR
ENTRY: 120.0
Va=120.0
Vb=120.0
Vab=208.0 Vbc=208.0 Vca=208.0
Ib=00.00
PHASE B
0.000
0.000
0.000
0.000
1.000
1.000
PHASE B
0.00
0.00
0.00
0.00
1.00
1.00
38
Vc=120.0
Ic=00.00
PHASE C
0.000
0.000
1.000
PHASE C
0.00
0.00
1.00