2.0 Power Supply Test/Current

Equipment Needed
Battery Current Test Fixture (P/N 9100-0690-TF),
Power Supply 12V 10A,
DMM (e.g. Agilent 34401A Digital Multimeter) or equivalent,
AED PLUS, Quick-Test, Test Fixture (P/N 9100-0567-TF).
Setup
Install the Battery Current Test Fixture (P/N 9100-0690-TF) into the battery well.
On the DMM, set Function to DC Amps by pressing Shift button then the DC I button.
Connect the RED lead from the I jack of the DMM to one of the Meter Leads on the Battery
Current Test Fixture (P/N 9100-0690-TF).
Connect the BLACK lead from the LO jack (located above the I jack) of the DMM to the
other Meter Leads on the Battery Current Test Fixture (P/N 9100-0690-TF).
Connect the power supply RED and BLACK leads, respectively into the Power Supply
posts on the Battery Current Test Fixture (P/N 9100-0690-TF).
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Set the Power Supply to 12.00V.
Test
2.1
2.2
2.3
2.4
2.5
1–4
T
ESTS
Action
Turn the Power Supply on.
Press the I Measure switch on
the Battery Current Test Fixture.
Press and release the On/Off
button.
Press the I Measure switch on
the Battery Current Test Fixture.
Press and hold the On/Off
button to power off the unit.
www.zoll.com
Expected Result(s)
Verify the unit issues a voice
and text prompt UNIT OK, and
the unit is turned off.
Observe the current
measurement on the Digital
Multimeter is between 10uA to
70uA.
Verify that the unit powers on in
rescue mode. Wait for the
following prompts are given:
ADULT PADS, DONT'T TOUCH
PATIENT ANALYZING... before
taking measurement.
Observe the current
measurement on the Digital
Multimeter is between 0.1A to
0.4A.
Verify unit displays UNIT
POWERING OFF message.
Pass
Fail
9650-0309-01 Rev. E