T10FH
8
Settings (Option 3, Code SU2)
8.1 Torque output signal
The factory setting for the frequency output voltage is 5 V (symmetrical, com
plementary RS422 signals). The frequency signal is on pin 4 opposite pin 1.
You can change the output voltage to 12 V (asymmetrical). To do this, change
switches S1 and S2 to position 1 (and pin 1 →
Fig. 8.1: Switch for changing the frequency output voltage
8.2 Setting up the zero point
In the case of the torque flange with the voltage output option (SU2), you can
access two potentiometers by removing the stator cover. You can use the zero
point potentiometer to correct certain zero point deviations. The adjustment
range is a minimum of "400 mV at nominal (rated) amplification. The end
point potentiometer is used for adjustment at the factory and is capped with
varnish so that it cannot be turned unintentionally.
CAUTION
Turning the end point potentiometer changes the factory calibration of
the voltage output.
A2037−4.0 en
).
5 V
symmetrical
asymmetrical
Pos. 2
S1
Pos. 1
S2
35
12 V
HBM