Setting Up Zero Point - HBM T10F Mounting Instructions

Torque flange
Hide thumbs Also See for T10F:
Table of Contents

Advertisement

T10F
Fig.5.1:
Solder bridges for shifting the frequency output voltage

5.4 Setting up zero point

If you have a torque measurement flange with the voltage output option, you
can remove the identification plate to gain access to two potentiometers. You
can use the zero point potentiometer to correct certain zero point variances.
The range of compensation is ™400mV. The end point potentiometer is used
for balancing at the factory and is capped with varnish so that it cannot be
turned unintentionally.
CAUTION
If you turn the end point potentiometer, you will change the factory cali-
bration.
Fig.5.2:
Setting up voltage output zero point
™5V
12V
asymmetrical
symmetrical
End point
Zero point
27