If the electrical connection of the KODAK MULTILOADER 700/700 PLUS SIDE-BY-SIDE
KIT /5000 RA is hardwired a MAIN POWER SWITCH consisting of a 2-pole switch with a
solid neutral and common trip must be-
•
located on a wall adjacent to the SIDE-BY-SIDE KIT/5000 RA
•
Easily accessible
•
visible from the front of the SIDE-BY-SIDE 5000 RA SYSTEM.
Ensure that the electrical connection to the mains for
•
KODAK MULTILOADER 700 is as stated in Site Specification N-704
•
KODAK MULTILOADER 700 PLUS is as stated in Site Specification 8B7005
•
KODAK X-OMAT 5000 RA PROCESSOR is as stated in Site Specification 5B6338
If the SIDE-BY-SIDE KIT /5000 RA or the ML700, ML700 PLUS or the PROCESSOR is
connected with a POWER CORD and PLUG to the mains, easily accessible and visible wall
outlets close to the SIDE-BY-SIDE KIT /5000 RA must be provided.
After the installation of the complete system check the electrical safety. Follow your local regulations.
This equipment includes parts and assemblies sensitive to damage from electrostatic discharge. Use
The unpacking and the assembly of the SIDE-BY-SIDE KIT cannot be done by one person. This is a
KODAK AG, Stuttgart
ELECTRICAL SAFETY
Ensure that all panels are mounted properly.
CAUTION
care to prevent damage during all service procedures.
two man job. The SBS is very heavy.
NOTE
1-3
II 3066-2
10/98