Cal/Verify Data

View the results of the latest calibration or verification. This includes the
following:
1.
2.
The display then shows the span gas calibration data:
The term span value on the display is the value the system read during the
calibrate/verify operation, and the term span gas is the setpoint value en-
tered using the Cal Gas Values option from the Calibrate key menu.
3.
The following then appears on the display:
5-10 | Thermox Series 2000 WDG Insitu Analyzer
O
span calibration gas value vs. O
2
Zero calibration gas value vs. zero calibration gas reading
Span calibration gas drift
Zero calibration gas drift
Span Gas and Zero Gas
Span Value or Zero Value
Span Drift or Zero Drift
Both the latest calibration and verification data will be stored in
memory.
NOTE
From the Calibrate key menu, select Cal/Verify Data.
Select Calibration Data or Verify Data, depending on whether you
want to see the latest calibration or verification results.
Press Enter to then view the zero gas values.
span calibration gas reading
2
Refer to the calibration gas values entered
using the Cal Gas Values menu option. These
values are fixed.
Value the system reads from the cal gas.
Difference between the entered calibration gas
value and the calibration gas values read by
the system.
Span Value: xx % O2
Span Gas: xx % O2
Span Drift: xx % O2
Zero Value: xx % O2
Zero Gas: xx % O2
Zero Drift: xx % O2