Ametek Thermox WDG Insitu User Manual page 22

With series 2000 control unit
Table of Contents

Advertisement

Sensor
Principle of
Operation
Output Range
Accuracy
Response
Drift
Linearity
Repeatability
Max. Flue Gas Temp / Probe Type Material
1250°F (677°C)
Sample Pressure
Environment
Ambient Temp.:
Relative Humidity: 0% to 90%, non-condensing
IEC Installation
Max Altitude::
Enclosure
UL Type (NEAM 4X) standard. Optional explosion-proof and purged versions available. Purged
version for sub-atmospheric process pressure only.
Power Requirements 115 VAC, ± 10%, 47-63 Hz, 150 VA max. 230 VAC, ± 10%, 47-63 Hz, 150 VA max.
Calibration Gas Requirements
Use calibration gases @ 10 PSIG, 3 to 6 SCFH (0.70 kg/cm
System Compliance
EMC Compliance:
Safety Compliance:
NOTES
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas.
2-2 | Thermox Series 2000 WDG Insitu Analyzer
Direct sensing insitu probe suitable for applications where the flue gas
temperature does not exceed 1250°F (677°C) and where combustibles
measurement is not required. Zirconium oxide for net oxygen mea-
surement.
From 0-1% to 0-100%
± 1% of measured value or ± 0.05% oxygen, whichever is greater
63% of a process step change < 20 secs.
< 0.1% of cell output per month (< 0.005% O
applied)
± 1% of output span
± 1% of output span
310 SS
± 10 inch water gauge
-5°F to 160°F (-20°C to 70°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Overvoltage Category II
IEC Pollution Degree 2
5000 meters
O
Span Gas: Air or from 1.0 to 100% O
2
O
Zero Gas: 2% or from 0.1 to 10% O
2
2004/108/EC
73/23/EEC
per month with 2% O
2
2
, 1.41 to 2.82 L/min.)
, balance N
2
2
, balance N
2
2
2