Set Current Mode; Output Filtering Option - Ametek Thermox WDG Insitu User Manual

With series 2000 control unit
Table of Contents

Advertisement

Set Current Mode

Define either a 0-to-20 or 4-to-20 mA analog output range for the selected
analog output port. This output is dependent on both the range of read-
ings you define and on the function you track on the analog output.
Setting the current mode
1.
Select an analog output port.
2.
From the Analog Range key, select Set Current Mode.
3.
Select 4-to-20 mA Mode or 0-to-20 mA Mode. Default selection is 4-to-
20 mA.
The following will then appear on the display:

Output Filtering Option

Filter out quick transitions in readings to allow a smoother analog output.
Setting output filtering
1.
Select an analog output port.
2.
From the Analog Range key, select Output Filtering. You will be
prompted to enter an output filtering value:
where XX is the current filter value.
100 equals no filtering, and 1 equals the highest possible filter value.
The default value is 100 (no filtering).
NOTE
3.
Type a number between 1 and 100, then press Enter. Note that you
should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc.
Out # y is set for
4 - 20 ma
o
r
Out # y is set for
0 - 20 ma
[100 =FAST
1 = SLOW]
Fltr = XX New #?
Controller / User Interface | 4-17