Analog Range Key - Ametek Thermox WDG Insitu User Manual

With series 2000 control unit
Table of Contents

Advertisement

Analog Range Key

Figure 4-3. Analog Range Key menu.
You can direct process readings to either of the two analog output ports
on the rear of the control unit. If you have the combustibles option, you
can also have process readings directed to two additional analog output
ports activated with this option. All functions in this chapter are directed
to a specific analog output port.
ANALOG KEY
Set Current Range
Set Track/Hold
Set Current Mode
Output Filtering
Select Function
ANALOG RANGE KEY
Set Current
Set Track/Hold
Set Current
Output Filtering
Select Function
FUNCTIONS/OPTIONS
Define what information should be sent to the analog
output port during a calibration. For example:
4% O
= 4 mA
2
10% O
= 20 mA
2
Hold output at last process reading during a cal or
verify, or track the calibration readings.
Define analog output range:
0 to 20 mA
4 to 20 mA
Smooth out variations in readings
Define the information the analog output should track:
% O
2
Cell temperature
Cell Millivolts
Thermocouple Millivolts
Range
Mode
Controller / User Interface | 4-13