Defining alarm set point
1.
From the Alarm menu, select Alarm Set Points.
Once you select the Alarm Set Points menu option, the system
alerts you that Alarms 1 and 2 are reserved for the watchdog and ser-
vice alarms, and are not software selectable. Press Enter to continue
NOTE
to the configurable alarms.
You will be prompted to define the Alarm 3 set point value (this option
won't appear if Alarm 3 is set to activate based on the start of a calibra-
tion or verification - see the "Select Function" section in this chapter for
details):
2.
Enter the Alarm 3 set point value, then press Enter (just press Enter if
you don't want to change the current value).
You are prompted to enter the Alarm 4 set point value:
3.
Enter the Alarm 4 set point value. Press Cancel or Enter to exit this
menu option.
Alarm #3= xx %O2
New Value?
Alarm #4= xx %O2
New Value?
Controller / User Interface | 4-21