4-16 | Thermox Series 2000 WDG Insitu Analyzer
Set Track/Hold Option
Define the signals to send to the selected analog output port during a cali-
bration or verification. You have the option to either hold the last process
reading or track calibration readings. Separate decisions can be made for
verify and calibrate operations, so you can hold the last process readings
during verifications, but track calibration gas readings during calibrations.
Setting track or hold
Select an analog output port.
From the Analog Range key, select Set Track/Hold.
You can choose to either hold process readings or track calibration
readings on the analog output port. Separate choices can be made for
calibration and verify operations.
For calibration operations, select one of the following:
Your choice will be confirmed on the display:
For verify operations, select one of the following:
Your choice will be confirmed on the display:
Hold during cal
or
Track during cal
Out # 1 will TRACK
during Calibration
or
Out # 1 will HOLD
during Calibration
Hold during Verify
or
Track during Verify
Out # 1 will TRACK
during Verify
or
Out # 1 will HOLD
during Verify