PARAMETER
O2 SPAN GAS DURATION
O2 ZERO GAS DURATION
COMB SPAN GAS DURATION
RECOVERY DURATION
K FACTOR
OUTPUT # 1 FUNCTION
CURRENT MODE,RANGE
TRACK/HOLD CAL VERIFY
OUTPUT FILTER # 1
OUTPUT # 2 FUNCTION
CURRENT MODE,RANGE
TRACK/HOLD CAL VERIFY
OUTPUT FILTER # 2
OUTPUT # 3 FUNCTION
CURRENT MODE,RANGE
TRACK/HOLD CAL VERIFY
OUTPUT FILTER # 3
OUTPUT # 4 FUNCTION
CURRENT MODE,RANGE
TRACK/HOLD CAL VERIFY
OUTPUT FILTER # 4
RELAY CONFIGURE
ALARM 1
ALARM 2
ALARM 3
ALARM 4
ALARM 5
ALARM 6
ALARM 7
ALARM 8
E-2 | Thermox Series 2000 WDG Insitu Analyzer
DEFAULT SETTINGS
2:00 MINUTES
2:00 MINUTES
2:00 MINUTES
4:00 MINUTES
1
% OXYGEN
4 TO 20 MA = 0-10
HOLD,TRACK
100
% OXYGEN
4 TO 20 MA = 0-25
TRACK,TRACK
100
COMBUSTIBLES
4 TO 20 MA = 0-2000PPM
HOLD,HOLD
100
COMBUSTIBLE
4 TO 20 MA = 0-2000PPM
TRACK,TRACK
100
DEENGERIZE RELAYS ON ALARM
WATCHDOG
SERVICE
10% O2 HIGH
1% O2 LOW
1000 PPM COMB
2000 PPM COMB
2% METHANE
3% METHANE
CUSTOMER SETTINGS