Ametek Thermox WDG Insitu User Manual page 145

With series 2000 control unit
Table of Contents

Advertisement

5D/93
Alr1_flg
5E/94
Alm_config
Alm_config
What process alarm is associated with
S2
0 – oxygen
oxygen, combustibles, or methane
1 – calibration
2 – verify
3 – in cal/verify
Configuration for oxygen alarms, high or
IQ
low.
If a bit is 1, that is the condition
that the indicated alarm will trip on.
bit 0 – high oxygen, process alarm
bit 1 – low oxygen, process alarm
bit 2 – high combustible, process alarm
bit 3 – high methane, process alarm
bit 4 – high oxygen, alarm #2
bit 5 – low oxygen, alarm #2
bit 6 – high combustible, alarm #2
bit 7 – high methane, alarm #2
bit 8 – high oxygen, alarm #3
bit 9 – low oxygen, alarm #3
bit 10 – high combustible, alarm #3
bit 11 – high methane, alarm #3
bit 12 – Future use
bit 13 –Future use
bit 14 – Future use
bit 15 – Future use
Configuration for oxygen alarms, high or
S2
low. If a bit is 1, that is the condition
that the indicated alarm will trip on.
bit 0 – Future use
bit 1 – Future use
bit 2 – Future use
bit 3 – Future use
bit 4 – Future use
bit 5 –Future use
bit 6 – Oxygen Alarm #3. 0=Low, 1=High
bit 7 – Oxygen Alarm #4. 0=Low, 1=High
bit 8 – Future use
bit 9 – Future use
bit 10 – Future use
bit 11 – Future use
bit 12 – Future use
bit 13 – Future use
bit 14 – Future use
bit 15 – Future use
R/W
IQ
R/W
IQ
IQ
IQ
IQ
IQ
IQ
IQ
IQ
IQ
IQ
IQ
IQ
R/W
Serial Communications
|
A-17