Select Function; Service Alarm - Ametek Thermox WDG Insitu User Manual

With series 2000 control unit
Table of Contents

Advertisement

Select Function

Select which function Alarm 3 should track.
% Oxygen
When the system starts a calibration
When the system starts a verification
When the system starts a calibration or a verification (dual function)
When tracking the start of a calibration or verification, the alarm will de-
activate once the calibration or verification has been completed, including
any recovery time you have specified.
Defining the select function
1.
From the Alarm menu, select Select Function.
2.
Select Alarm3= Oxygen, Alarm3= In Cal., Alarm3= In Verify, or
Alarm3= Cal/Verify, depending on whether Alarm 3 should be associ-
ated with % oxygen levels, or with the start of a calibration or verifi-
cation, or with the start of a calibration or verification, or both (dual
function).
The control unit automatically exits from this menu once you make your
selection.

Service Alarm

The Service Alarm operates as normally closed (service alarm condi-
tion causes open contacts).
NOTE
The Service Alarm provides an alarm to alert you of possible system prob-
lems that require attention. Events that can trigger the service alarm to
activate are:
Error condition (detected by software)
System operational variance (such as an over-temperature or under-
temperature condition)
Calibration failure
Verification failure
Control unit power failure
Controller / User Interface | 4-23